16. Mass Spectrometry Basics for Proteomics¶
What this covers¶
This chapter is drawn from a guest lecture on mass-spectrometry (MS) based proteomics — the physical measurement underlying the "-omics" data the rest of this course analyses statistically. It answers two questions: how does a mass spectrometer turn a mixture of peptides into a spectrum, and why is the intensity it reports a noisy, non-linear stand-in for "how much peptide was there"? It closes by introducing peptide fragmentation and the naming convention used to read a sequence back out of a fragment spectrum. It assumes only that a protein is a chain of amino-acid residues joined end to end; no prior mass-spectrometry knowledge is assumed.
Why look past the sequence at all¶
Genome sequencing gives a protein's primary structure — its sequence of residues (the lecture's
example: ...YSFVATAER...) — but the sequence alone does not fix what the molecule does in a cell.
Proteomics is framed here as filling in the layers a sequence cannot show: secondary structure
(local structural elements), tertiary structure (the folded 3D shape), and two further layers that
are dynamic rather than fixed by the gene — modifications such as phosphorylation, which switch
function on and off, and processing: proteolytic events that target or activate a protein,
illustrated with trypsin and with platelet activity. Measuring the protein directly, at the level
mass spectrometry operates on, is how these layers become visible.
Amino acids and the peptide bond¶
Amino acids differ considerably in their physico-chemical properties — size, charge, hydrophobicity — and a protein backbone is built by joining them one at a time, losing a water molecule between the amino group of one residue and the carboxyl group of the next:
Repeating this reaction gives a chain with a free amino group at one end (the N-terminus) and a free carboxyl group at the other (the C-terminus), with each residue's distinguishing side chain (\(R_1, R_2, \dots\)) hanging off the backbone:
This directionality — a chain with a chemically distinct start and end — is what later lets a fragment ion be labelled by which end of the peptide it came from.
Anatomy of a mass spectrometer¶
A mass spectrometer is described as three functional parts plus a digitizer:
- The ion source gets (some of) the sample's molecules into the gas phase and charged — without a charge, an analyte has no \(m/z\) for the instrument to measure at all.
- One or more mass analyzers manipulate these gas-phase ions with electromagnetic fields to separate them by mass-to-charge ratio, \(m/z\).
- The detector records the arrival of ions.
- The digitizer (analog-to-digital converter) turns the detector's continuous analog signal into the discrete, plotted spectrum.
Ionizing peptides: MALDI and ESI¶
Two ion sources dominate proteomics, and they hand the mass analyzer very different populations of ions.
MALDI (Matrix-Assisted Laser Desorption/Ionization) mixes the analyte with a light-absorbing matrix and fires a pulsed UV laser at it (typically nitrogen, \(\nu = 337\text{ nm}\)); the resulting desorption and proton transfer from matrix to analyte produce ions in the gas phase under high vacuum. Peptides ionised this way are typically singly charged, and different analytes compete for the available charge — competitive ionisation, part of why intensity is not simply proportional to amount (see below). The name was coined by Karas and Hillenkamp (Anal. Chem., 1985); Koichi Tanaka received the 2002 Nobel Prize in Chemistry for demonstrating that MALDI could ionise biological macromolecules (Rapid Commun. Mass Spectrom., 1988).
ESI (Electrospray Ionization) instead nebulises the sample from a needle held at 3–5 kV and heated to roughly \(40\)–\(100^{\circ}\text{C}\); the resulting charged droplets evaporate, and either undergo charge-driven fission or expel ions directly toward the analyzer inlet. ESI typically produces multiply charged peptide ions (\(2^+\), \(3^+\), \(4^+\)). John Fenn shared the 2002 Nobel Prize in Chemistry for demonstrating ESI on biological macromolecules (Science, 1989); the lecture adds that the same physical process is used in fine-control thrusters on satellites and interstellar probes.
Resolving and detecting ions¶
Resolution is defined as the width of a peak at a given height — most often the width at 50% of peak height, the full width at half maximum (FWHM); an alternative definition uses the percent-valley height between two neighbouring peaks instead. Resolving power matters because a peptide does not produce a single peak: its isotopes spread the signal over several peaks close in \(m/z\), so the same peptide has both a monoisotopic mass and a (slightly higher) average mass depending on how much of that isotope envelope the instrument can resolve into distinct peaks versus blur into one.
Detection in most instruments is by electron multiplier: a single incident ion striking the first of several increasingly-charged dynodes triggers a cascade, each dynode boosting the electron count (the lecture's example steps dynode voltages from \(20\,\text{V}\) up to \(120\,\text{V}\)) until a single incoming ion produces on the order of \(10^6\) electrons out.
Why an intensity is not an amount¶
The quantification principle is simple to state: detector signal relates to the quantity of analyte present. Turning that into a usable number means, first, making each sample distinguishable — either by introducing a mass difference between samples so they can be run together, or by running each sample through the instrument as a separate experiment — then measuring, for each analyte, the intensity of its signal in each sample, and finally processing the resulting numbers statistically. The lecture illustrates this with mixtures combined at known ratios, such as \(1{:}2\), \(1{:}1\) and \(2{:}1\) — exactly the kind of comparison behind the error sources below.
Several things make the middle step, measuring intensity, harder than it sounds.
Intensity is not comparable across peptides. Different peptides ionise with different efficiency, so one peptide's MS1 intensity says nothing about another's abundance; a given peptide's intensity is informative only relative to that same peptide's intensity in another run or sample.
The detector response saturates. Plotting the measured \(\log_2\) ratio between two mixed samples against the ratio they were actually mixed at (the expected \(\log_2\) ratio), the response is linear near \(1{:}1\) but levels off as the true ratio becomes more extreme: the detector stops faithfully tracking intensity once it is far enough from the middle of its dynamic range (Gevaert, Proteomics, 2007).
Error grows with the ratio. Separately from that saturation, the percentage error on a measured ratio between two peaks increases the further the ratio departs from \(1{:}1\) (Vaudel, Proteomics, 2010) — and both effects, the lecture notes, remain clearly visible on modern Orbitrap instruments, not only older ones.
Peak-picking adds its own error. Before a "peak" exists as a number at all, the raw digitized signal has to be converted into a list of (\(m/z\), intensity) values — a step that is instrument-specific, sets the practical lower limit of the dynamic range (the signal-to-noise ratio), and by itself often contributes a 5–10% error to the final ratios. That conversion has to assume a peak shape: comparing peak-picking under three different mass windows (\(0.02\), \(0.04\), \(0.08\ \text{Da}\)) against a mismatched, "non-adapted" shape model shows the wrong shape alone adding on the order of another 10% error (Vaudel, Proteomics, 2010). Several software tools exist for this step (the lecture names Decon2LS), and — its own phrase — "there's actually more to a peak than just \(m/z\)", illustrated with the OpenMS TOPPView viewer, though the slide does not spell out what that additional information is.
The sample degrades before it is even measured. Serum proteins continue to be degraded over time after collection, regardless of the sampling tube used (Yi, J. Proteome Res., 2007) — a reminder that pre-analytical handling, not just instrument physics, is a source of error.
Modifications are more pervasive than expected. An open modification search (ionbot,
re-analysing the data behind Kim et al., Nature, 2014) finds a striking number of modified forms
even among six well-studied, highly abundant glycolytic enzymes — between 50 and 278 distinct
modified peptides per protein, and well over 100 distinct modifications on several of them
(glyceraldehyde-3-phosphate dehydrogenase, 166; pyruvate kinase PKM, 139; fructose-bisphosphate
aldolase A, 122; alpha-enolase, 121; triosephosphate isomerase, 117; phosphoglycerate kinase, 111),
spanning modification types from carbamylation and oxidation to succinylation and ammonia loss. A
search that only looks for the handful of modifications it was told to expect will miss most of
this.
From spectrum to sequence: MS/MS and fragment ions¶
Identifying which peptide produced a spectrum relies on fragmenting it and reading the fragment masses. Tandem MS does this with two mass analyzers used one after another (a single ion trap can also do the whole job by itself): the first analyzer acts as an ion selector, passing only ions at one chosen \(m/z\) — the precursor; fragmentation is then triggered; and the second analyzer records the \(m/z\) of the resulting fragments, which the detector reads out as before. The pipeline from the earlier figure gains an extra stage: source, then ion selector, then fragmentation, then fragment mass analyzer, then detector.
Because the peptide backbone has a definite direction — N-terminus to C-terminus, from the section above — a fragment ion is named by which terminus it retained and by how many residues it carries from that terminus: fragments that keep the N-terminus are called \(a\), \(b\) or \(c\) ions, and fragments that keep the C-terminus are called \(x\), \(y\) or \(z\) ions.
Internal fragments — pieces that have lost both termini — can also appear and be interpreted, but they don't fit this ladder scheme and are harder to use for reading a sequence directly off the spectrum. This naming was introduced by Roepstorff and Fohlmann (Biomed. Mass Spectrom., 1984) and by Klaus Biemann (Biomed. Environ. Mass Spectrom., 1988), and is commonly called the Biemann nomenclature — the choice of letters deliberately follows the Roman alphabet in each direction. In the ideal case, a full series of these ions — a complete ion ladder — lets the peptide sequence be read directly off the gaps between consecutive peaks, each gap being the mass of one residue.
The lecture's own material breaks off at exactly this point: the next slide promises to show what that ideal ladder looks like, but only a single extracted label ("L") survives from it, and the topics the deck's own table of contents promises after this one — database search algorithms, sequential search algorithms, decoys and false discovery rate calculation, and protein inference — are not present in the text this chapter was built from (see Sources).
Sources¶
All of this chapter's content comes from a single guest lecture, "Science meets life: MS-based
proteomics data analysis" by Lennart Martens (Compomics, Ghent University / VIB), delivered within
the statOmics SGA21 course:
docs/omics-statistics/statomics/sga21/docs/martens_proteomics_bioinformatics.md, converted from
martens_proteomics_bioinformatics.pdf (statOmics/SGA21 repository, commit
0ad787d4cc2bb2f4636440840a8a923cf6c09839), licensed CC BY-NC-SA 4.0.
The source PDF had no extractable text layer, so it was reconstructed by a model reading the pages; its own banner flags the prose as a paraphrase in places and every equation as unverified — the same caveat carries over to the two chemical equations reproduced here (the peptide-bond reaction and the backbone formula). No slide deck, transcript or exercise set was supplied separately for this lecture — the reconstructed markdown is the entirety of the source material, and every section above follows its slide-by-slide ordering. Figures extracted from the PDF are collected at the end of the source file by page number, without captions tying them to a topic, and are not reproduced here.
Named sources whose own content is not part of this chapter's material, only the claim they were cited for:
- the NCBI Science Primer, for the levels-of-structure framing;
- Karas and Hillenkamp, Anal. Chem. (1985), and Tanaka, Rapid Commun. Mass Spectrom. (1988), for MALDI;
- Fenn, Science (1989), for ESI;
- Eidhammer, Flikka, Martens and Mikalsen (Wiley, 2007), for mass resolution;
- Roepstorff and Fohlmann, Biomed. Mass Spectrom. (1984), and Biemann, Biomed. Environ. Mass Spectrom. (1988), for fragment-ion nomenclature;
- Gevaert, Proteomics (2007), and Vaudel, Proteomics (2010), for the quantification-error figures;
- Yi, J. Proteome Res. (2007), for serum protein degradation; and
- Kim et al., Nature (2014), as the dataset behind the
ionbot(https://ionbot.cloud) open modification search.